Accuracy Of Results

The accuracy of the Fatmeter depends upon the fat/oil content of the sample being measured and ranges from an uncertainty in the fat content of (+/-) 1% (95% confidence interval) at low levels to (+/-) 4% at very high levels (greater than 45%).

Therefore it is advisable that the operator follows the measurement methodology rigorously, and always takes readings as described in wall charts supplied.

Accuracy Guide

Fat/Oil Reading Accuracy
2 to 15% from ± 0.5% rising to ± 1.0% (13 – 15% at 14%)
16 to 30% from ± 1.0% rising to ± 2.0% (28 – 32% at 30%)
31% and above from > ± 2.0% rising to ± 4.0% (46 – 54% at 50%)

The results gained will allow the operator to take swift decisions regarding feeding or processing of the fish species being measured.

Other methods currently in use for determining the lipid content of fish samples are destructive and slow, eg. In fish, a section or fillet of the fish is taken and analysed for fat/oil content by weighing, and using solvents to extract the fat, or, by driving off the water content by heating. These methods, unlike the Distell Fatmeter, obviously cannot be used on live fish.