The accuracy of the FATMETER depends upon the fat/oil content of the sample being measured and ranges from an uncertainty in the fat content of (+/-) 1% (95% confidence interval) at low levels to (+/-) 4% at very high levels (greater than 45%).
Therefore it is advisable that the operator follows the measurement methodology rigorously, and always takes readings as described in wall charts supplied.
Accuracy Guide
| Fat/Oil Reading |
Accuracy |
| 2 to 15% |
from.. ± 0.5% rising to.. ± 1.0% (13 - 15% at 14%) |
| 16 to 30% |
from.. ± 1.0% rising to.. ± 2.0% (28 - 32% at 30%) |
| 31% and above |
from.. > ± 2.0% rising to.. ± 4.0% (46 - 54% at 50%) |
The results gained will allow the operator to take swift decisions regarding processing of the type of meat being measured.
|